Complete blood count (CBC)

Medonic M32

Clinical performance for a complete blood count (CBC)

A comparison of performance of Medonic M32 with reference instrument for complete blood count (CBC) analysis.

Medonic M32 automated hematology analyzer is routinely used in laboratory diagnostics for determination of patients’ blood status. This work demonstrates the performance of Medonic M32 3-part hematology analyzer in comparison with a more technically advanced 5-part reference analyzer in complete blood count (CBC) analyses of patient samples taken from the normal routine screening. The results show that the analyzers are in good agreement, indicating the suitability for use of Medonic M32 in general health screenings.

Introduction

A CBC is highly useful in general screenings as a tool to aid in diagnosis and monitoring of disease conditions. Automated instruments for this type of analyses were developed as early as in the 1950s. The Medonic systems were introduced by Ingemar Berndtsson and Bram Bottema, founders of Medonic AB in 1982 (now part of Boule Diagnostics) and both with a long history and experience in hematology, clinical chemistry, and blood-banking engineering.

Before, blood cell counts were performed manually by microscopy. Although manual examination of blood smears is still used as a control method for verification of results from abnormal samples, the automated hematology analyzers have largely replaced the manual method for determination of hematology parameters in the routine use.

The Medonic M32 system is an automated hematology analyzer for in vitro diagnostic use under laboratory conditions (Figure 1). The analyzer is intended for determination of hemoglobin (HGB) concentration, for counting of red blood cells (RBC) and platelets (PLT) as well as for counting and differentiation of white blood cells (WBC) into three subpopulations, namely lymphocytes (LYM), mid-sized white cells (MID, mainly monocytes), and granulocytes (GRAN, mainly neutrophils, eosinophils and basophils). The measurement principles of the Medonic M32 are based on impedance for cell counts and spectrophotometry for HGB.

Although such a 3-part hematology analyzer provides enough information for the smaller local hospital laboratory, trends show an increased interest in 5-part instruments, typically used in larger central hospital and hematology laboratories, also for use in small physician office laboratories (POL).

While a 5-part analyzer offers improved WBC assessment, differentiating them into neutrophils (NEU), lymphocytes (LYM), monocytes (MONO), eosinophils (EOS), and basophils (BASO), a 3-part instrument can offer great cost benefits to general screenings of patients’ blood status (1).

The objective of this study was to evaluate the performance of Medonic M32 3-part hematology analyzer against a 5-part reference instrument.

medonic-family (3)

Fig 1. Medonic M32 automated 3-part hematology analyzer is available in four versions. While both M32B and M32M support open tubeaspiration, M32M features an integrated mixer. M32C and M32S support closed-tube sampling to minimize the risks associated with contaminated blood. In addition, M32S is equipped with an Auto Loader for up to 2 × 20 samples – just load and walk away

Materials and methods

The following material was used in this study:

  • Medonic M-series M32C Hematology Analyzer
  • Medonic M-series Diluent
  • Medonic M-series Lyse
  • Boule Con-Diff Normal
  • Boule Con-Diff Low
  • Boule Con-Diff High

The evaluation of Medonic M32 (test instrument) was performed in collaboration with a Swedish hospital against Sysmex™ XE-5000 (reference instrument) in accordance with the standard SS-EN 13612 for compliance with the demands in the European IVD directive (98/79/EC). The correlation studies were based on > 340 samples taken from the normal routine. After analysis on the reference instrument, samples were run in the open tube (OT) inlet on the Medonic M32 system. All samples were analyzed in single assays.

Results

Results are presented in scatter plots in Figure 2, with the 95% confidence interval (prediction interval) of the indicated linear regression. As shown from the summary given in Table 1, the performance of Medonic M32 fulfills the specifications, and can thus be considered in good agreement with reference instrument.

Figure_02
Fig 2. Correlation of (A) RBC, (B) PLT, (C), and WBC, as well as (D) LYM, (E) MID and (F) GRAN for unflagged samples, between test system and reference systems.

Table 1. Summary of results

Conclusion

This study demonstrates that the performance of Medonic M32 hematology analyzer is in good agreement with that of the reference instrument. Although manual microscopic examination of blood smears is recommended as complementary method for confirming analytical data, the results indicate the suitability of Medonic M32 for use in routine hematology analysis.